8440 Central Avenue.
Suite # 2C, Newark
CA: 94560, USA

Phone: (510) 791-0951
Fax: (510) 794-4330
www.transducertech.com

 
P Series Hydrogen Sulfide (H2S) Sensor – PH2S1000

 

Mechanical Specifications

Dimensions 1.39″ diameter ´ 0.58″ height
Weight Approximately 10 gram
Material Polypropylene
Electrolyte Etching liquid
Connections Contact pads
Gas Supply Diffusion through porous membrane
 
Nominal Range 0 to 200 PPM
Maximum Overload 1000 PPM
Measuring Principle Electrochemical Oxidation of H2S
Working Electrode Potential +300 mV with respect to the reference electrode (Consult factory for other usage.)
Output Signal, Zero, 25degC 0 to ± 0.75 mA
Output Signal, Span, 25degC 0.30 to ± 0.15 mA / PPM
Operating Temperature -20 to 50o C (0 – 35o C recommended)
Response Time (10% to 90%) < 60 s typical at 20o C
Output Linearity Linear
Stabilization time 30 minutes when first installed in a circuit
Operating Pressure Range ± 0.2 atm
Operating Humidity Range 10 to 95% RH Continuous (0 to 100% intermittent)
Estimated Service Life 24 month  (Depends on application)
Storage Temperature Range 4 to 18o C Recommended
Storage Pressure Range ± 0.2 atm
Storage Humidity Range 60 to 80% RH Recommended without sealed package
Storage Life 1 years in sealed package
Lower Detection Limit 2 PPM depends on circuitry
Resolution ± 2 PPM depends on circuitry
Warranty One year (extended warranty available)
Contact TTI for application information. TTI reserves the right to alter design features and specifications without notice.

 

© 2006 Transducer Technology, Inc. CareersPrivacy Policy